Frye FONIX 7000 Manual do Utilizador

Consulte online ou descarregue Manual do Utilizador para Saúde e higiene Frye FONIX 7000. Frye FONIX 7000 User Manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir

Resumo do Conteúdo

Página 1 - Hearing Aid Test System

Hearing Aid Test SystemOperator’s Manual Version 1.51©2006, Frye Electronics, Inc. All rights reservedOct. 23, 2006FONIX® 7000

Página 2

2 FONIX 7000 Hearing Aid Analyzerwere implemented and tested on other Frye instruments. The FP40 analyzer introduced “function buttons” that adapt to

Página 3 - Contents

92 FONIX 7000 Hearing Aid AnalyzerIf there is a sequence of tests that you would like to perform frequently, you can save it into the 7000 Test Syste

Página 4

Basic Sound Chamber Tests 933.8.2 Loading a Saved Test SequenceIt is not necessary to permanently save a custom test sequence, but doing so allows yo

Página 5

94 FONIX 7000 Hearing Aid Analyzer1. If desired, follow the instructions in Section 3.8.2 to load a test sequence. You can do this by following the

Página 6

Basic Sound Chamber Tests 95Measurement Settings• Coupler Type: The type of coupler being used in making sound chamber measurements. Selections of C

Página 7

96 FONIX 7000 Hearing Aid Analyzer• Misc Meas Delay: The amount of time each subsequent tone is presented before the measurements is made in one of

Página 8

Chapter 4 97 Advanced Coupler Tests4.1 Enhanced DSPThe Enhanced DSP measurement screen was designed to give you more infor-mation about digital hear

Página 9 - Introduction & Setup

98 FONIX 7000 Hearing Aid AnalyzerTechnical detailsThe digital processing delay measurement is taken by sending a short impulse from the sound chambe

Página 10 - 1.3 Options

Automated Test Sequences 99It then continues through a complete cycle and terminates when it reaches the baseline again. This signal, although consis

Página 11

100 FONIX 7000 Hearing Aid AnalyzerWhen measuring a set of binaural hearing aids, both the phase and the delay graphs should match fairly closely. 4.

Página 12 - 1.4 Accessories

Automated Test Sequences 101Figure 4.2—Battery Current Test1. mA/Freq graph2. mA/Ampl graph3. Static test (estimated battery life)4. Selected tes

Página 13

Introduction and Setup 3with a probe microphone. Using real-ear measurements makes it possible to individualize the fitting of a hearing aid since a

Página 14 - 1.4.2 Real-Ear Accessories

102 FONIX 7000 Hearing Aid Analyzer4.2.2 mA/FreqThe mA/Freq test shows how much battery current the hearing aid drains when exposed to different fre

Página 15

Automated Test Sequences 103• You can view the numerical data results by entering the local menu by pressing [MENU] and setting DATA/GRAPH to DATA b

Página 16 - 1.4.3 Optional Accessories

104 FONIX 7000 Hearing Aid AnalyzerA few technical details about this test:• Choose between showing test results in SPL (Output) and Gain in the loc

Página 17

Automated Test Sequences 105Figure 4.4—Attack & Release Test1. Attack graph2. Release graph3. Measured Attack & Release times and test se

Página 18

106 FONIX 7000 Hearing Aid AnalyzerFor most hearing aids, the most interesting part of the attack and release mea-surements occurs at the beginning o

Página 19 - 1.5.1 Front Panel Layout

Chapter 5 107 Automated Test SequencesThere are several automated coupler test sequences available for the 7000 hear-ing aid analyzer. The test seque

Página 20

108 FONIX 7000 Hearing Aid AnalyzerHere are the main differences between ANSI 87 and ANSI 96:• AGC aids are tested at reduced reference test gain in

Página 21 - T630 mAL

Automated Test Sequences 1093. Position the hearing aid/coupler assembly in the test chamber with the microphone of the aid at the reference point o

Página 22 - 1.5.3 Safety

110 FONIX 7000 Hearing Aid Analyzer3. For ANSI 87 and ANSI 96, set AGC aids to achieve the greatest possible compression or as otherwise specified b

Página 23

Automated Test Sequences 1112. Push [START] to continue the test.3A. If you chose to perform a telecoil measurement, do the following steps when the

Página 24 - 1.6 Setup

4 FONIX 7000 Hearing Aid AnalyzerHA-2, 2-cc Coupler Dimensions per ANSI S3.7 for testing behind-the-ear aids, eyeglass aids, and body aids.Coupler M

Página 25

112 FONIX 7000 Hearing Aid Analyzer5.1.5 Running an ANSI 87 TestAfter leveling the sound chamber as described in Section 5.1.1, setting up the aid a

Página 26 - Probe/Reference mics

Automated Test Sequences 113Figure 5.1.6—ANSI 03 test results, Frequency Response Screen 9. RTG Measured: Measured reference test gain10. EQ Inp Noi

Página 27 - 1.7 Maintenance

114 FONIX 7000 Hearing Aid Analyzer18. I/O curve selection box: the key to the I/O graph, listing which frequencies go with which curve.19. Attack an

Página 28

Automated Test Sequences 1151. Run the ANSI test sequence.2. Press [F6].3. Use [∨] to select Store Crvs Coup, if necessary. 4. Press [>] to st

Página 29

116 FONIX 7000 Hearing Aid Analyzer• OSPL90 Sweep Meas Delay—The delay between frequencies in the OSPL90 sweep.• Res Crv Sweep Start Delay—The dela

Página 30 - 1.7.5 Warranty

Automated Test Sequences 1175.2 ANSI S3.42-1992 (ANSI 92)The ANSI Option includes a test sequence based on the ANSI S3.42-1992 stan-dard. This stand

Página 31 - General Operation

118 FONIX 7000 Hearing Aid AnalyzerFor some aids, the analyzer will pause the test and tell you to turn the aid down to reference test gain. Adjust t

Página 32

Automated Test Sequences 119Figure 5.2.3—ANSI 92 test results5.2.4 Viewing ANSI measurements in the Coupler screenIt’s possible to copy ANSI respons

Página 33 - General Operation 25

120 FONIX 7000 Hearing Aid Analyzer5.2.5 Explaining the Menu SettingsThere are only a couple of menu settings in the local menu of the ANSI 92 scree

Página 34

Automated Test Sequences 1215. Level the sound chamber, if necessary (see Section 5.1.1).6. Plug the battery simulator into the jack near the batte

Página 35 - 2.1.4 Using the local menus

Introduction and Setup 5Microphone Adapter 14 mm to 1" (25.4) mm adapter for coupling the M1950E microphone and reference microphone to a 1-inch

Página 36 - 2.2 Setup Menu

122 FONIX 7000 Hearing Aid Analyzer7. Reference test gain measured (actual)8. Reference frequency9. Total harmonic distortion10. Equivalent input

Página 37

Automated Test Sequences 123• Misc. Start Delay—The delay after the first tone is presented and before the first measurement is taken in the HFA/SPA

Página 38 - 2.2.4 Explaining the settings

124 FONIX 7000 Hearing Aid Analyzer5.4.2 Viewing JIS ResultsFigure 5.4.2See Figure 5.4.2 for the following explanation of test results.1. OSPL90 re

Página 39

Automated Test Sequences 1255.4.3 Viewing JIS measurements in the Coupler screenIt is possible to copy JIS response curves to the Coupler screen whe

Página 40

126 FONIX 7000 Hearing Aid AnalyzerJIS Settings• Equiv Noise Test—Status of the EIN measurement. Turns it ON or OFF. Misc. Settings• Print Label—St

Página 41 - 2.3 Source Types

Chapter 6 127Real-Ear Measurements6.1 IntroductionThe Real-Ear Option gives you the capability to make probe microphone mea-surements inside your cl

Página 42 - 2.3.1.1 Pure-tone settings

128 FONIX 7000 Hearing Aid AnalyzerThese screens are:• Audiogram Entry screen [F1]• Target screen [F2]• Insertion Gain screen [F3]• SPL screen [F

Página 43 - 2.3.1.3 Delay times

Real-Ear Measurements 12912"Speaker(45 degrees)Reference MicProbe Mic12"Reference MicProbe MicTop ViewFront ViewSpeaker(45 degrees)Figure 6

Página 44 - 2.3.1.5 Pure-tone Filter

130 FONIX 7000 Hearing Aid Analyzer6.2.3 Inserting the probe tubeThere are several different methods used for properly inserting the probe tube. Here

Página 45

Real-Ear Measurements 131WARNINGChoose OUTPUT LIMITING carefully (see procedure below). You don’t want to damage your clients’ hearing or to cau

Página 46 - 2.3.2.1 Noise reduction

6 FONIX 7000 Hearing Aid AnalyzerLoudspeaker Swing Arm This swing arm may be substituted for the speaker floor stand & shelf. It is a convenien

Página 47 - 2.3.2.3 Composite filter

132 FONIX 7000 Hearing Aid AnalyzerSANITATION NOTICEDO NOT REUSE PROBE TUBES.Use a new probe tube for each ear to prevent the possible spr

Página 48 - 2.3.2.4 Composite type

Real-Ear Measurements 1333. Press [F5] to select the Composite signal as the source type.4. Press [START] to start a measurement. 5. Insert the pr

Página 49 - Figure 2.3.2.5

134 FONIX 7000 Hearing Aid AnalyzerFigure 6.2.4—Leveling the sound field speaker 1. Follow the instructions found in Sections 6.2.1 and 6.2.2 to pos

Página 50 - 2.4 Printing

Real-Ear Measurements 135If the instrument didn’t successfully level, there are several things you can try. • Check the position of the sound field

Página 51 - 2.4.2 Adding a label

136 FONIX 7000 Hearing Aid Analyzer4. Make sure HTL is selected above F2. You should see a cursor in the HTL column for the selected ear in the midd

Página 52

Real-Ear Measurements 1376.3.2 Deleting Audiometric InformationIf you make a mistake, it’s often useful to be able to delete all or part of the input

Página 53

138 FONIX 7000 Hearing Aid Analyzer6.3.4 Entering Bone DataThe NAL-NL1 target uses bone data in its calculations. If the patient’s bone thresholds ar

Página 54

Real-Ear Measurements 139To perform the coupler measurement:1. Insert the 50 ohm insert earphone into the jack on the back of the 7000 test system l

Página 55

140 FONIX 7000 Hearing Aid Analyzer5. Press [F1] from the Real-Ear Navigation screen to enter the Audiogram Entry screen.6. Press [MENU] from the A

Página 56

Real-Ear Measurements 141Figure 6.3.5E—Taking an RECD Measurement6.3.6 Modifying a TargetOnce you input an audiogram in the Audiogram Entry screen, a

Página 57 - 2.5 Display & Data

Introduction and Setup 7Monitor Headset Plugs into the remote module, and allows the user to listen to the signal being measured by the probe micro-p

Página 58

142 FONIX 7000 Hearing Aid Analyzer• REDD: The real-ear to dial difference (only selectable when REDD is set to CUSTOM)• TARG IG: The mid-level tar

Página 59 - 2.5.2 Viewing numerical data

Real-Ear Measurements 1436.3.7 Entering an REDD transformThe REDD transform affects how dB HL values are converted to dB SPL. To enter a measured RED

Página 60 - 2.7 Remote Module

144 FONIX 7000 Hearing Aid AnalyzerIn other words, the bottom graph of the insertion gain screen shows the differ-ence between the sound field measur

Página 61 - 2.8 Leveling

Real-Ear Measurements 1456.4.2 Measuring in the Real-ear SPL screenIn the past ten years, the SPL method of performing real-ear measurement has incre

Página 62

146 FONIX 7000 Hearing Aid Analyzerthis section apply to measurements performed in both screens. Any differences between operations in the two screen

Página 63 - Basic Sound Chamber Tests

Real-Ear Measurements 1475. Enter the Insertion Gain or Real-ear SPL screen by using [F3] or [F4] in the Real-ear Navigation screen, respectively.6.

Página 64

148 FONIX 7000 Hearing Aid Analyzerthe local menu. The average unaided response will now be displayed on the measurement graph.To auto-adjust the una

Página 65 - 3.2 The Leveling Process

Real-Ear Measurements 149pop-up selection menu. Use the up-down arrow keys to make the desired selection. Press the right arrow key to complete the s

Página 66 - 3.3 Hearing Aid Setup

150 FONIX 7000 Hearing Aid Analyzeranother measurement curve. It is recommended to perform real-ear measure-ments as 50, 65, 80, and 90 dB SPL. For t

Página 67 - MICROPHONE

Real-Ear Measurements 1516.4.6 Using Semi-Auto and Fully-Auto ModeThe Real-ear Insertion Gain and Real-ear SPL screens have basic implementa-tions of

Página 68

8 FONIX 7000 Hearing Aid AnalyzerCIC Coupler Non-standard (0.4 cc) coupler used for realistic testing of CIC hearing aids. This coupler is always use

Página 69

152 FONIX 7000 Hearing Aid Analyzer1. Enter the Real-ear Insertion Gain or Real-ear SPL screen.2. Press [MENU] to enter the local menu.3. Select S

Página 70 - 3.3.4 Open-Ear Hearing Aids

Real-Ear Measurements 153To view curve data in the Real-ear Insertion Gain screen:1. Enter the Real-ear Insertion Gain:2. Press [MENU] to enter the

Página 71

154 FONIX 7000 Hearing Aid Analyzer4. Use [<, >] to select DATA.5. Press [EXIT] to return to the Real-ear SPL screen. The numerical data dis-

Página 72 - 3.3.5 Body Aids

Real-Ear Measurements 155• From 3000 Hz–8000 Hz, measured data points from seven consecutive frequencies 100 Hz apart are averaged to get the curve

Página 73 - 3.3.6 Eyeglass Aids

156 FONIX 7000 Hearing Aid Analyzer6.4.12 Using Output LimitingWhen performing real-ear measurements, it is very important not to expose your client

Página 74

Real-Ear Measurements 1577. Press [∨] repeatedly until the chosen source is OFF. 8. Press [START] to start the measurement.9. Instruct your client

Página 75

158 FONIX 7000 Hearing Aid Analyzer6.5.2 Choosing a speech weightingBy default, the Digital Speech (and Composite) signal uses the ANSI S3.42 speech

Página 76 - 3.4 Basic Measurements

Real-Ear Measurements 1592. Select the Digital Speech signal source with [F5].3. Press [MENU] to enter the local menu.4. Select Bias Tone under So

Página 77 - 3.4.2 Running a Test Curve

160 FONIX 7000 Hearing Aid Analyzer5. Press [START] to measure the frequency response of the hearing aid. Press [STOP] when the measurement has stab

Página 78 - 3.4.3 Example

Real-Ear Measurements 1614. Press [LEVEL] to level the sound field speaker.5. Use [F2] to select REUG 1. (Press [F2] to open the pop-up selection m

Página 79

Introduction and Setup 9Open Ear Coupler Non-standard coupler used for realistic testing of open ear hearing aids.#5 battery pill Provides battery c

Página 80

162 FONIX 7000 Hearing Aid Analyzer14. Press [START] to start the frequency response measurement. (The “reverse” measurement will reappear when the m

Página 81

Real-Ear Measurements 1637. Test as usual. With the reference microphone disabled, it is even more criti-cal than usual that the patient not move af

Página 82

164 FONIX 7000 Hearing Aid AnalyzerFigure 6.8—Visible Speech screen before a measurement is taken. The shaded regions are below and above the patient

Página 83

Real-Ear Measurements 1656.8.2 Viewing the Real-time Visible Speech DisplayDuring the Visible Speech test, the analyzer displays four changing measur

Página 84

166 FONIX 7000 Hearing Aid AnalyzerFigure 6.8.3—Completed Visible Speech test.6.8.4 Measuring multiple envelopesThe clinician can measure up to three

Página 85 - 3.4.9 Battery Current Drain

Real-Ear Measurements 1676.9 CROS/BICROS AidsWhen performing real-ear measurements on a Body aid, CROS, or Bi-CROS aid, we suggest the following setu

Página 86 - 3.4.10 Telecoil

168 FONIX 7000 Hearing Aid Analyzer6.9.1 Measuring the Head-Baffle EffectThe measurements described in this section are for both CROS and BI-CROS hea

Página 87

Real-Ear Measurements 1696.9.2 Measuring the Overcoming of the Head-Baffle EffectThis section describes how to measure how well the hearing aid overc

Página 88

170 FONIX 7000 Hearing Aid Analyzer6.9.2.2 BI-CROSUnaided—Measurement of “better” ear canal (baffled by head)Set up the 7000 test system as follows.•

Página 89

Real-Ear Measurements 1716.9.3.1 CROSUnaided—Measurement of “good” earSet up the 7000 test system as follows.• Reference microphone ON• Sound field L

Página 90

10 FONIX 7000 Hearing Aid AnalyzerSound Chamber Spring Damper Factory-installed.or Assembly Kit. This spring assembly fits onto the back of the soun

Página 91 - Basic Sound Chamber Tests 83

172 FONIX 7000 Hearing Aid Analyzer6.9.4 Measuring Insertion Loss to the “Good” Ear (CROS)The method for measuring insertion loss in the patient’s “g

Página 92 - 3.5 Digital hearing aids

Appendix A 173 SpecificationsACOUSTIC DRIVE SIGNALFrequencies: Composite: 200 to 8000 Hz in 100 Hz intervals SINE: 200 to 8000 Hz 1/12 oct, nearest

Página 93

174 FONIX 7000 Hearing Aid AnalyzerSPECTRUM MODE AVERAGING Spectral : off, 2,4,8 and 16.HARMONIC DISTORTION ANALYZERType: Selectable for 2nd, 3rd,

Página 94

Specifications 175ELECTRONICS MODULEColor: Blue module case, white front panel and gray bezelSize: 17.4"W x 6.5"H x 14.6"D (44.2

Página 95

176 FONIX 7000 Hearing Aid AnalyzerEar Level Adapter: (044-1007-00) Per requirements of ANSI S3.22-1996 for adapting ear level aids to the HA-2 c

Página 96 - 3.6 CIC Feature

Appendix B 177 CalibrationB.1 Calibrating the microphonesTo calibrate the 7000 Test System’s microphones, you will need a sound level calibrator such

Página 97

178 FONIX 7000 Hearing Aid Analyzer6. If necessary, adjust the gain of the 1950E microphone by using the flathead screwdriver to adjust the “gain” p

Página 98

Calibration 179Figure B.1.C—Adjusting the reference pot with a Phillips-head screwdriver.5. Remove the reference microphone from the adapter and tur

Página 99

180 FONIX 7000 Hearing Aid Analyzer10. Turn on the sound level calibrator and look at the “Probe Mic” level in the Microphone Calibration window. It

Página 100

Calibration 181Figre B.2—Calibrating the insert earphone

Página 101

Introduction and Setup 111.5 Layout, Controls, and SafetyThis section describes the basic layout of the front and rear panel of the 7000 test system,

Página 102

182 FONIX 7000 Hearing Aid Analyzer

Página 103

Appendix C 183 Troubleshooting GuideThese are the most common problems that typically cause instrument failure. Please check these troubleshooting su

Página 104

184 FONIX 7000 Hearing Aid AnalyzerCoupler Problems:1. HA-2 Coupler (BTE) adapter tubing is missing, loose, or cracked Replace with #13 thickwall tu

Página 105 - Advanced Coupler Tests

Appendix D 185 The FONIX CIC FeatureBackgroundZwislocki built an ear simulator coupler years ago to better approximate the real ear’s impedance varia

Página 106 - 4.1.2 Signal Phase

186 FONIX 7000 Hearing Aid Analyzera good hearing aid fitting. It also felt that the approach taken in the use of the MZ coupler has been well accept

Página 107 - 4.1.3 Procedure

The FONIX CIC Feature 187Knowing the physical volume of the occluded ear canal and its frequency response variations, it is possible to calculate the

Página 108 - 4.2 Battery Current Test

188 FONIX 7000 Hearing Aid AnalyzerCIC CORRECTION FACTORSdB Hz dB Hz dB Hz dB Hz-8.7 200 -0.95 2200 3.2 4200 4.54 6200-8.6 300 -0.5 2300 3

Página 109

Appendix E 189 Fitting Formula TablesThe conversions used by the FONIX 7000 Test System to convert audiograms to target gain curves are the following

Página 110 - 4.2.3 mA/Ampl

190 FONIX 7000 Hearing Aid Analyzer POGO Prescription of Gain/Output (POGO) for Hearing Aidsby: Geary A. McCandless, PhD, & Poul Erik Lyregaard

Página 111 - 4.3 Coupler I/O

Fitting Formula Tables 1911/3, 1/2, 2/3 GainThe 1/3-2/3 Insertion Gain Hearing Aid Selection Guide by: Robert LibbyHearing Instruments vol 37 #3, 19

Página 113

12 FONIX 7000 Hearing Aid Analyzer• To navigate from screen to screen• To toggle a selection• To pop-up up a selection menuFor a more complete des

Página 114

192 FONIX 7000 Hearing Aid Analyzer

Página 115 - Automated Test Sequences

Appendix F 193 Probe SPL Mode DescriptionThe Target IG is converted to the Target SPL in the following steps:(Conversion tables appear on the followi

Página 116 - 5.1.1 Leveling for ANSI

194 FONIX 7000 Hearing Aid AnalyzerTABLE 1 Average Real-Ear Unaided Response (REUR) FREQ GAIN FREQ GAIN FREQ GAIN FREQ GAIN(Hz) dB (Hz) dB (Hz) d

Página 117

Probe SPL Mode Description 195 TABLE 3 HL to real-ear SPL (or SPL to HL) conversion table from ANSI S3.6-1989 Table G.1 FREQ (Hz) dB 250 19.0

Página 118

196 FONIX 7000 Hearing Aid Analyzer

Página 119

Appendix G 197 Glossary of TermsAGC Automatic Gain Control (or Automatic Volume Control). A feature on a hearing aid that automatically adjusts the

Página 120 - 5.1.6 Viewing ANSI Results

198 FONIX 7000 Hearing Aid AnalyzerBICROS A CROS aid that has two microphones, one at each ear, sending both signals to one ear.BLOOMING A

Página 121

Glossary of Terms for the FONIX 7000 199DATA In FONIX analyzers, the word “data” refers to displaying curve information as a table of numbers.

Página 122

200 FONIX 7000 Hearing Aid AnalyzerGIFROC The inverse of “CORFIG.” GIFROC is the transformation that, when added to a coupler response, will give th

Página 123

Glossary of Terms for the FONIX 7000 201INTER- Distortion generated by the faulty mixture of different input frequencies. MODULATION This is

Página 124

Introduction and Setup 131.5.2 Rear Panel LayoutThe rear panel of the 7000 test system contains most of the external connec-tions for the analyzer. S

Página 125 - 5.2.2 Running ANSI 92

202 FONIX 7000 Hearing Aid AnalyzerPOWER The output of an aid, in dB SPL.PROBE A miniature microphone with a thin, flexible tube that is

Página 126

Glossary of Terms for the FONIX 7000 203RMS Root Mean Square. An overall measurement of the signal going in or coming out of a hearing aid. Wh

Página 127

204 FONIX 7000 Hearing Aid AnalyzerWARBLE TONE A pure tone with slight, but rapid, frequency variations.WEIGHTED A measure of the amplification

Página 128 - 5.3 IEC

205IndexAAdaptive AGC 36AGC frequency 116AIDED DATA 153Aided response 148Aid type 95ANSI 84, 158ANSI 03 107, 108, 110ANSI 92 117ANSI Option 3

Página 129 - 5.3.3 Viewing IEC Results

206 FONIX 7000 Hearing Aid AnalyzerEEar-level adapter 4, 59Ear hook 7, 129Enhanced DSP 97Equiv Noise Test 115, 122, 126EXIT 12, 28Eyeglass 65FF

Página 130

Index 207Noise reduction 35, 38, 88Noise Red (Comp) 95Noise Red (Tone) 95, 115, 122, 125Noise suppression 84, 86, 88, 159NORMAL 33, 69OOccluded

Página 131 - 5.4 JIS

208 FONIX 7000 Hearing Aid AnalyzerSpecifications 173Speech mapping. See live speech mappingSpeech weighting 158SPL 145, 147, 149, 150, 193START

Página 132 - 5.4.2 Viewing JIS Results

14 FONIX 7000 Hearing Aid Analyzer11. PROBE: Connects the main module to the remote module (used for real-ear measurements). KEYBD: Not currently

Página 133

Introduction and Setup 15Connection of peripheral equipment to the 7000Compliance with IEC 60601-1-1 Safety requirements for medical electrical sys-t

Página 134

16 FONIX 7000 Hearing Aid AnalyzerOtherwise, please report all safety related concerns to the Frye factory: Frye Electronics, Inc. 9826 SW Tigard

Página 135 - Real-Ear Measurements

Introduction and Setup 171.6.2 Locating and arranging the 7000 test systemLocate the 7000 test system in a moderately quiet area, such as a private o

Página 136 - 6.2 Real-Ear Setup

18 FONIX 7000 Hearing Aid AnalyzerPRINTERSOUND CHAMBEREARPHONESPKRMONITORRS232AUXKEYBDMICROPHONEGAINSCOPEPROBEFONIX 7000 Serial No. 1234Manufactured

Página 137 - Front View

Introduction and Setup 194. Depending on whether you have a floor stand or a swing arm, either (1) Unfold and extend the floor stand to the desired

Página 138

20 FONIX 7000 Hearing Aid AnalyzerThe 7000 microphones should be wiped with a dry cloth. Excess moisture may damage the microphones.Solvents and abra

Página 139 - Real-Ear Measurements 131

Introduction and Setup 21Figure 1.7.3A—Serial number locationThe software version of your 7000 test system can be found on the Opening screen. See Fi

Página 140 - SANITATION NOTICE

ContentsChapter 1: Introduction & Setup1.1 History. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 141

22 FONIX 7000 Hearing Aid Analyzer1.7.4 Contacting the FactoryIf you have any questions regarding your instrument, have your serial number and softwa

Página 142

Chapter 2 23 General OperationThis chapter describes the general operation of the 7000 test system, including basic operation of the buttons, screen

Página 143 - 6.3 Real-Ear Targets

24 FONIX 7000 Hearing Aid AnalyzerThere are three basic uses of the function keys:• To navigate from screen to screen• To toggle a selection• To p

Página 144

General Operation 252.1.2 Navigating through the screensThe function keys are used to move from screen to screen. Here is a flow chart of all the ava

Página 145 - 6.3.3 Creating a Target

26 FONIX 7000 Hearing Aid Analyzer2.1.3 Using the pop-up help windowsEach operational screen on the 7000 test system has a local help window designed

Página 146 - 6.3.4 Entering Bone Data

General Operation 272.1.4 Using the local menusPressing the [MENU] button will pop up a local menu that, like the help win-dow, will be overlaid on y

Página 147 - Figure 6.3.5A Figure 6.3.5B

28 FONIX 7000 Hearing Aid Analyzer2.1.5 Using the EXIT and RESET buttonsThe [EXIT] and [RESET] buttons can be used to exit from any screen, menu, or

Página 148 - to insert microphone

General Operation 292.2.1 Saving and loading default settingsAll of the settings in the Setup menu (described in Section 2.2.4) can be saved as defau

Página 149 - 6.3.6 Modifying a Target

30 FONIX 7000 Hearing Aid Analyzer2.2.3 Setting the user modeThere are two different user modes available on the 7000 test system: Beginner and Exper

Página 150

General Operation 31START SCREEN: The screen shown when the analyzer is powered on or the [RESET] button is pressed. COUPLER sets this screen to the

Página 151 - 6.4 Real-Ear Measurements

2.3.1.2 Noise reduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 152

32 FONIX 7000 Hearing Aid AnalyzerNOISE RED (TONE): Pure-tone noise reduction for real-ear measurements. See Section 2.3.1.2.NOISE RED (COMP): Compos

Página 153

General Operation 33harmonic distortion measurement should be ignored. Choose ON to apply this rule. Choose OFF to ignore this rule.ANSI OPTION: Modi

Página 154

34 FONIX 7000 Hearing Aid Analyzer2.3.1.1 Pure-tone settings There are several different settings available for pure-tone sweeps. You can change all

Página 155

General Operation 35HFA (High Frequency Average) 1000, 1600, 2500 SPA (Special Purpose Average) 800, 1250, 2000 SPA 1250, 2000, 3150 SPA 1600,

Página 156

36 FONIX 7000 Hearing Aid AnalyzerIn addition to the delays, the aid type will also have an effect on the way some measurements are made. With a sele

Página 157

General Operation 37tone sweeps with speech-weighted Composite or Digital Speech signals. For this reason, we have made it possible to “weight” the r

Página 158

38 FONIX 7000 Hearing Aid AnalyzerA significant advantage of using a composite signal is that each frequency pres-ent in the signal can be individual

Página 159

General Operation 39Figure 2.3.2.2—Example of IM distortion2.3.2.3 Composite filterThere are two main types of composite filters (also known as “spee

Página 160

40 FONIX 7000 Hearing Aid AnalyzerFigure 2.3.2.3—Comparison of the ICRA shape and the ANSI shape2.3.2.4 Composite typeThere are two types of the trad

Página 161

General Operation 41Chirp Composite signal is the standard composite signal used on the FP40 and FP35 portable hearing aid test systems. Although the

Página 162

3.4.2 Running a Test Curve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 163

42 FONIX 7000 Hearing Aid Analyzer2.3.2.6 Digital Speech technical detailsThe digital speech signal works by turning the composite signal on for a se

Página 164 - 6.4.12 Using Output Limiting

General Operation 43To set the default printer, and/or to set the external printer language and color choice, use the Setup menu:1. Press [MENU] fro

Página 165 - 6.5 Digital Hearing Aids

44 FONIX 7000 Hearing Aid AnalyzerIf you want to include a label on all printouts by default:1. Press [MENU] from the Opening screen to enter the Se

Página 166

General Operation 45• Init failed 0: No memory available for the converted screen buffer• Init failed 1: No memory available for the label bitmap•

Página 167

46 FONIX 7000 Hearing Aid AnalyzerFigure 2.4.4A—The internal printerStep II: Loading paper into the printer1. Turn on the instrument.2. Place the H

Página 168 - 6.6 Directional Hearing Aids

General Operation 47Figure 2.4.4C—Loading paper into the printer4. While holding the paper in place behind the rubber roller, manually turn the thum

Página 169

48 FONIX 7000 Hearing Aid Analyzer8. Tear off excess paper by pulling the paper against the thermal head at approximately a 45º angle.9. Close the

Página 170 - 6.7 Open Ear Hearing Aids

General Operation 492.4.7 Putting multiple screens on a pageWhen printing with an external printer, it is possible to save paper by putting two scree

Página 171 - 6.8 Visible Speech

50 FONIX 7000 Hearing Aid AnalyzerCurve Name: This identifies the name of each curve. The first letter is derived from its measurement type—coupler (

Página 172

General Operation 51RMS Out: The root-mean-square of the measured curve.Correction Factors (COR): The coupler correction factors used to make the mea

Página 173

5.1.7 Testing Digital Hearing Aids. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 174

52 FONIX 7000 Hearing Aid AnalyzerFigure 2.5.2—Data display of a composite measurement2.6 Computer ConnectionThe 7000 test system is equipped with an

Página 175 - 6.9 CROS/BICROS Aids

General Operation 53Figure 2.7—Remote Module2.8 LevelingLeveling is the process by which the response of the sound chamber or sound field is measured

Página 176 - PROBE MIC

54 FONIX 7000 Hearing Aid Analyzer• WEAK_LEVELED: Tolerances are the same as SEMI-LEVELED but the max-imum output for the speaker is less than 100 d

Página 177 - 6.9.2.1 CROS

Chapter 3 55 Basic Sound Chamber TestsIn the Coupler screen, you can view measurement curves in dB SPL or dB Gain, and run them with Composite, Digit

Página 178 - 6.9.2.2 BI-CROS

56 FONIX 7000 Hearing Aid AnalyzerRefer to Figure 3.1.1 for the following explanation of the graphical display:1. Type of display. This will be eith

Página 179 - 6.9.3.2 BI-CROS

Basic Sound Chamber Tests 574. Signal type, speech weighting, leveling status, and bias status (Digital Speech only).5. Selected noise reduction.6.

Página 180 - REFERENCE MIC

58 FONIX 7000 Hearing Aid AnalyzerFigure 3.2.1—Leveling the sound chamberNote: See Section 5.1.1 for the ANSI standardized leveling procedures.3.2.2

Página 181 - BATTERY CURRENT MEASUREMENT

Basic Sound Chamber Tests 593.3.1 Explaining the Coupler Types The HA-1 direct access coupler is used to test in-the-ear and canal aids, and aids wit

Página 182

60 FONIX 7000 Hearing Aid Analyzerto duplicate that used by the manufacturer to obtain the original data. For ANSI tests, use #13 thick-walled tubing

Página 183 - Specifications 175

Basic Sound Chamber Tests 614. Seal any vent on the aid with a small kernel of Fun-Tak.5. Complete the acoustical sealing of the aid to the coupler b

Página 184 - GUARANTEE

6.4.6 Using Semi-Auto and Fully-Auto Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 185 - Calibration

62 FONIX 7000 Hearing Aid Analyzer3.3.3 Behind-The-Ear (BTE) Aids1. Insert the coupler microphone into the HA-2 2-cc coupler. 2. Couple the aid to

Página 186

Basic Sound Chamber Tests 63We have also developed a non-standard coupler for open-ear hearing aids. It has two design purposes: 1) To make an conven

Página 187

64 FONIX 7000 Hearing Aid AnalyzerFigure 3.3.4D—Comparison of couplers with open-ear hearing aid. Curve 1 was measured using the Open Ear coupler, Cu

Página 188

Basic Sound Chamber Tests 65 Figure 3.3.5A—Coupler setup Figure 3.3.5B—Body Aid in test chamber 3.3.6 Eyeglass Aids1. Follow the manufacturer’s

Página 189 - Calibration 181

66 FONIX 7000 Hearing Aid AnalyzerFigure 3.3.6—Eyeglass Aid Set-up3.3.7 Wireless CROS and BICROS AidsThe key to testing these aids is to get the prop

Página 190

Basic Sound Chamber Tests 67Figure 3.3.7—Testing Wireless CROS Hearing Aids4. Place the receiver on top of the lid of the test chamber. The total de

Página 191 - Troubleshooting Guide

68 FONIX 7000 Hearing Aid AnalyzerAdditional Notes for Testing Wireless Instruments1. There must not be any metal objects between the transmitter an

Página 192 - Printer Problems:

Basic Sound Chamber Tests 693.4.1 Signal TypesThe following signal types are available:DIGITAL SPEECH—A randomly interrupted composite signal designe

Página 193 - The FONIX CIC Feature

70 FONIX 7000 Hearing Aid AnalyzerFigure 3.4.2—Pure-tone measurement3.4.3 ExampleFollowing the instructions above, run 4 Gain curves using amplitudes

Página 194

Basic Sound Chamber Tests 716. Use [F2] to select CRV 2.7. Use [∧] to select a source level of 60 dB SPL.8. Press [START] to run the second measur

Página 196

72 FONIX 7000 Hearing Aid AnalyzerFigure 3.4.3B—Family of curves. Notice how the curves show less gain as the amplitude increases. This aid shows a g

Página 197 - Fitting Formula Tables

Basic Sound Chamber Tests 733.4.5 Running a Single Frequency MeasurementAt times, you may want to find out what the frequency response of the hearing

Página 198

74 FONIX 7000 Hearing Aid Analyzer3. Use [<, >] to choose AVERAGE.4. Use [∨, ∧] to select Avg Freqs under Aid Settings.5. Use [<, >] t

Página 199 - 1/3, 1/2, 2/3 Gain

Basic Sound Chamber Tests 753RD: Distortion present at the third harmonics, or three times the presented frequency.TOTAL: Distortion present at the s

Página 200

76 FONIX 7000 Hearing Aid Analyzer3.4.8 Intermodulation DistortionThe composite signal is helpful for identifying intermodulation distortion (IM), a

Página 201 - Probe SPL Mode Description

Basic Sound Chamber Tests 773.4.9 Battery Current DrainYou can measure the battery current drain in the Coupler screen if you use battery pills (inst

Página 202 - TABLE 2

78 FONIX 7000 Hearing Aid Analyzer3.4.10 TelecoilTelecoil measurements can be performed in the Coupler screen using any of the source signals availab

Página 203 - TABLE 4

Basic Sound Chamber Tests 798. Use [∧, ∨] to change the intensity of the magnetic field. The choices are: OFF, 1.00, 1.78, 3.16, 5.62, 10.0, 17.8, 3

Página 204

80 FONIX 7000 Hearing Aid Analyzeris in a vertical position. If necessary, use Fun Tak to help hold the aid in that position. See Figure 3.4.10B.5.

Página 205 - Glossary of Terms

Basic Sound Chamber Tests 81Speaker PlacementThe speaker in the 7000 sound chamber is positioned on the right side of the chamber at a 45º angle, as

Página 206

Chapter 1 1Introduction & SetupWelcome to the newest innovation in hearing aid testing! The 7000 Test System is the long-awaited successor to the

Página 207

82 FONIX 7000 Hearing Aid AnalyzerFigure 3.4.11B—Forward directional measurement.Reverse measurementNext, open the sound chamber and position the hea

Página 208

Basic Sound Chamber Tests 83Figure 3.4.11C—Reverse directional measurementFigure 3.4.11D—Directional measurement screenshot

Página 209

84 FONIX 7000 Hearing Aid Analyzer3.5 Digital hearing aidsAll digital aids can be tested, but some of the high-end models require a little more thoug

Página 210

Basic Sound Chamber Tests 855. Use [F5] to choose the DIG SPEECH signal source. It will now have the ICRA speech weighting. See Figure 3.5.2 for a c

Página 211

86 FONIX 7000 Hearing Aid AnalyzerFigure 3.5.3—Using the bias signal with Digital Speech. CRV 1-3 have bias signals at 500 Hz, 1000 Hz, and 4000 Hz r

Página 212

Basic Sound Chamber Tests 872. Use [F2] to select CRV 1. (Press [F2]. Use [∨, ∧] to select CRV 1. Use [>] to complete the selection and close the

Página 213

88 FONIX 7000 Hearing Aid Analyzer3.5.5 Understanding TerminologyWith the new testing terminology of hearing aid analyzers combined with the new tech

Página 214

Basic Sound Chamber Tests 898. Press [START] to perform the measurement. You should see CIC appear in the Curve Characteristics box in the “Cor” col

Página 215 - Index 207

90 FONIX 7000 Hearing Aid AnalyzerTo use the OES Option:1. Attach the MZ coupler to the aid. See Table 3.7 for details on choosing the correct MZ co

Página 216

Basic Sound Chamber Tests 91Figure 3.7—Comparison of the responses with the MZ and HA-2 couplers. Curve 2 was measured with an MZ coupler and OES co

Comentários a estes Manuais

Sem comentários