Frye FONIX FP40D Manual do Utilizador

Consulte online ou descarregue Manual do Utilizador para Saúde e higiene Frye FONIX FP40D. Frye FONIX FP40D User Manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 164
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 0
FONIX
®
FP40/FP40D
PORTABLE HEARING AID ANALYZER
OPERATOR’S MANUAL
Software Version 3.70 © September 2005 Frye Electronics, Inc.
Rev. Jun. 04, 2007 All Rights Reserved.
A Note on this Manual
The instructions in this manual are for software version
3.70 and above, with references to earlier software.
However, you may contact Frye Electronics for a more
appropriate manual if you have earlier software.
Vista de página 0
1 2 3 4 5 6 ... 163 164

Resumo do Conteúdo

Página 1 - FP40/FP40D

FONIX® FP40/FP40D PORTABLE HEARING AID ANALYZEROPERATOR’S MANUALSoftware Version 3.70 © September 2005 Frye Electronics, Inc. Rev. Jun. 04, 2007

Página 2

2 FONIX FP40 Portable Hearing Aid Analyzer Also in 1994, we introduced the FP40-D desk model analyzer. This unit is always equipped with the real e

Página 3 - Table of Contents

92 FONIX FP40 Portable Hearing Aid Analyzer 5.4.3 Viewing the SPL screenSee Figure 5.4.3 for a picture of the SPL testing screen.1. Curve box con

Página 4

Real-Ear Measurements 93 2. Set up the client, insert the probe tube, and level the sound field speaker as described in Section 5.1.3. Insert th

Página 5

94 FONIX FP40 Portable Hearing Aid Analyzer Figure 5.4.4—SPL screen displayed on an external monitor5.5 Audibility Index (AI)The Audibility Index

Página 6

Real-Ear Measurements 95 5.5.1 Viewing the AI displaySee Figure 5.5.1 for a picture of the Audibility Index display.1. dB HL graph for unaided a

Página 7 - Appendices

96 FONIX FP40 Portable Hearing Aid Analyzer 13. Output limit status14. Selected ear15. Graph containing unaided and aided gain response5.5.2 Perfo

Página 8

Real-Ear Measurements 97 5.6.1.1 To perform the coupler measurementThe coupler measurement part of the RECD is saved into the analyzer’s permanen

Página 9 - Chapter 1: Introduction

98 FONIX FP40 Portable Hearing Aid Analyzer 8. Plug a 50-ohm insert earphone into the external speaker jack on the back of the FP40. An adapter ma

Página 10 - 1.3 Features & Options

Real-Ear Measurements 99 3AINSERTEARPHONELapel clipMicrophone jackExternal speaker jackProbe micProbe micFoam eartip3AINSERTEARPHONE3AINSERTEARPHO

Página 11 - 1.3.4 Battery Pack Option

100 FONIX FP40 Portable Hearing Aid Analyzer Figure 5.6.1.2B—Measured RECD5.6.2 Performing coupler measurements to a DSL target1. Enter your clie

Página 12 - 1.3.6 RS232 Option

Real-Ear Measurements 101 Figure 5.6.2A—Coupler gain measurement with a DSL target11. Press F5 to enter SSPL MODE. You should see a series of star

Página 13 - 1.4 Accessories

Introduction 3 The use of the Composite or Digital Speech signals can uncover the presence of intermodulation distortion in a hearing aid. Intermo

Página 14 - 1.4.2 Optional Accessories

102 FONIX FP40 Portable Hearing Aid Analyzer If the FP40 has the OES Option, it is also possible to make a prescription that uses the modified Zwis

Página 15

Real-Ear Measurements 103 5.7.2 Taking the FOG measurement1. Follow the instructions found in Section 5.2 to generate a real-ear target.2. Pres

Página 16 - 1.4.3 Real-Ear Accessories

104 FONIX FP40 Portable Hearing Aid Analyzer 7. Age of client.8. Instructions for performing the SSPL 90 measurement.Note: Once a measurement has

Página 17

Real-Ear Measurements 105 5.7.5 Checking an aid against a prescriptionThis procedure allows the operator to reenter data from the client’s initia

Página 18 - 1.5 Layout & Controls

106 FONIX FP40 Portable Hearing Aid Analyzer 5.7.6 Accounting for venting effectsYou can make 2-cc targets more accurate by modifying the target t

Página 19 - 1.5.3 Front Panel Lamps

Real-Ear Measurements 107 5.8 MiscellaneousThis section describes a few real-ear features accessible from the Setup Menu.5.8.1 Single frequency

Página 20 - 1.5.5 Rear Panel Controls

108 FONIX FP40 Portable Hearing Aid Analyzer 5.8.3 Reset LevelThe reset level is the sound pressure level that the analyzer automatically returns

Página 21 - Paper Release Lever

Real-Ear Measurements 109 5.9 Body, CROS, and BI-CROS aidsWhen using the FP40 probe with a Body aid, CROS, or BI-CROS aid, we suggest the followi

Página 22 - 1.6 FP40 Setup

110 FONIX FP40 Portable Hearing Aid Analyzer 5.9.2 Testing CROS and BICROS aidsFour Goals: 5.9.2.1 Measure the head baffle effect 5.9.2.2 Meas

Página 23 - 1.7 Miscellaneous

Real-Ear Measurements 111 5.9.2.2 How Well the Aid Overcomes the Head-Baffle EffectCROSA. Unaided—Measurement of “good” ear canal (baffled by hea

Página 24 - 1.7.4 Warranty

4 FONIX FP40 Portable Hearing Aid Analyzer 1.3.5 ID OptionThe ID Option personalizes the printout strips with the owner’s name and address or phon

Página 25 - Chapter 2: General Operation

112 FONIX FP40 Portable Hearing Aid Analyzer The difference curve, labeled “Insertion Gain” on the screen, shows the benefit of adding the second m

Página 26 - 2.2.2 Saving changes

Real-Ear Measurements 113 B. Aided—Measurement of “better” earSame setup as A, except…• Complete aid in place in better ear and set at use gain•

Página 27 - 2.3 Using Function Keys

114 FONIX FP40 Portable Hearing Aid Analyzer 5.10 FM SystemsFor users who test FM Systems, a comprehensive guidebook is available free of charge.

Página 28

Real-Ear Measurements 115 9. Press START/STOP to start the measurement. While the measurement is running, you may want to adjust the angle of the

Página 30 - 2.4.1.2 Noise Reduction

Spectrum Analysis 117 Chapter 6: Spectrum Analysis6.1 Spectrum ModeThis mode is only available on FP40 instruments that have the Real-Time/Composit

Página 31 - 2.4.2.1 ICRA vs. ANSI

118 FONIX FP40 Portable Hearing Aid Analyzer 6. Use F3 to select AIDED 1.7. Place the hearing aid in the ear. The aid should be turned off.8.

Página 32 - 2.4.2.2 Noise Reduction

Spectrum Analysis 119 APPLICATION 3: REAL EAR AIDED RESPONSE (REAR) IN SPL USING AN EXTERNAL SOUND SOURCE. This application is similar to the Insert

Página 34 - 2.6 Battery Current Drain

Telecoil Testing 121 Chapter 7: Telecoil TestingTelecoil testing is available on serial number 940000 and above, manufactured in September, 1994 and

Página 35

Introduction 5 HA-1 2-cc Coupler Dimensions per requirements of ANSI S3.7 for testing in-the-ear aids.HA-2 2-cc Coupler Dimensions per requirement

Página 36 - 2.7 Using the DATA button

122 FONIX FP40 Portable Hearing Aid Analyzer 7. Press F1 to return to the Main Coupler screen.8. Press F2 to turn ON the telecoil feature.9. Ori

Página 37 - 2.9 Printing

Telecoil Testing 123 4. Use the AMPLITUDE and FREQUENCY knobs to select MAIN F2 under FUNCTION KEY DEFIN.5. Press START/STOP repeatedly to select T

Página 38

124 FONIX FP40 Portable Hearing Aid Analyzer 7.4 Testing1. Use the amplitude knob to increase the level of the magnetic field. If you are using t

Página 39 - 2.9.4 Printing a label

Specifications 125 Appendix A: SpecificationsSINE SIGNALFrequencies 1/12 octave frequencies from 200 to 8000 Hz, closest 100 Hz,Normal Sweep w

Página 40 - 2.10 The Opening Screen

126 FONIX FP40 Portable Hearing Aid Analyzer Tolerance +/- 0.01V, no load. Battery Type Selection Under software control from front panel, bu

Página 41

Specifications 127 Auto Shutdown General shutdown after no operation of controls for 15 minutes (battery operation only).Battery Charger Buil

Página 42

128 FONIX FP40 Portable Hearing Aid Analyzer PHYSICAL DESCRIPTIONDimensions 20.125” x 14.750” x 6.5” (50,5 x 36,9 x 16,25 cm) (with lid on case).

Página 43 - 3.1 The Main Coupler Screen

Calibration 129 Appendix B: CalibrationCalibrationYour new instrument has been calibrated at the factory. However, from time to time, you will w

Página 44 - 3.2 Leveling

130 FONIX FP40 Portable Hearing Aid Analyzer Calibrating the Probe Microphone1. Place the probe calibration adapter in the 1” to 14 mm adapter as

Página 45

Calibration 131 7. PUSH F6 COMPEN (compensate) Probe. When the signal stops, the probe has been compensated and the compensation has been s

Página 46 - 3.3 Hearing Aid Setup

6 FONIX FP40 Portable Hearing Aid Analyzer FM40 Microphone Provided if the Real-Ear Option is not ordered.Battery Pills (With 12" [30 cm] cabl

Página 48

History of Changes in the FP40 133 Appendix C: History of ChangesVERSION 1.0 (ORIGINAL RELEASE 3/90)VERSION 1.2 (5/90)Added: JIS Standard.VERSIO

Página 50 - 3.4 Frequency responses

Custom RECD Test 135 135 Appendix D: Custom RECD TestThere are two methods of getting Custom RECD measurements with the FP40 using the Probe Option

Página 51 - 3.4.2 Taking the measurement

136 FONIX FP40 Hearing Aid Analyzer Part C: The RECD Calculation.On a separate sheet of paper, subtract the REAR measurements (Part B) from the 2cc c

Página 52

The FONIX CIC Option 137 BackgroundZwislocki built an ear simulator coupler years ago to better approximate the real ear’s impedance vari-ation wi

Página 53

138 FONIX FP40 Portable Hearing Aid Analyzer of the cavity between the hearing aid and the TM by moving the aid closer to it, we should expect to s

Página 54

The FONIX CIC Option 139 dB Hz dB Hz dB Hz dB HzCIC CORRECTION FACTORS-8.7 200 -8.6 300 -8.7 400-8.75 500-9 600-8.9 700-8.8

Página 56

Storage Compartments in the FP40 141 Appendix F: Storage Compartments in the FP40pull up hereAccessories are stored in three compartments located

Página 57 - 3.6 Directional Hearing Aids

Introduction 7 6040 Sound Chamber For control of external noises.Open Ear Coupler Non-standard coupler used for realistic testing of open ear hea

Página 59

FONIX FP40 Portable Hearing Aid Analyzer 143Appendix G: Troubleshooting GuideThese are the most common problems that typically cause instrument fa

Página 60 - 3.7 The CIC Option

FONIX FP40 Portable Hearing Aid Analyzer 144 FONIX FP40 Portable Hearing Aid Analyzer3. A bump or peak in the low frequency response curve a.

Página 61 - 3.8 The OES Option

FONIX FP40 Portable Hearing Aid Analyzer 145Appendix H: Probe SPL Mode DescriptionThe Target IG is converted to the Target SPL in the following st

Página 62

FONIX FP40 Portable Hearing Aid Analyzer 146 FONIX FP40 Portable Hearing Aid AnalyzerTABLE 1 Average Real-Ear Unaided Response (REUR) FREQ GA

Página 63 - 4.1 ANSI S3.22-2003

FONIX FP40 Portable Hearing Aid Analyzer 147TABLE 3 HL to real-ear SPL conversion table from ANSI S3.6-1989 Table G.1 FREQ (Hz) dB 250 19.0

Página 64

FONIX FP40 Portable Hearing Aid Analyzer

Página 65

FONIX FP40 Portable Hearing Aid Analyzer 149Appendix I: DSL Programming NotesNotes on the DSL® ProgrammingThe DSL® (Desired Sensation Level) sof

Página 66 - 4.1.4 Viewing the results

FONIX FP40 Portable Hearing Aid Analyzer 150channel instrument can meet those targets. If the compression ratio targets are very different in dif

Página 67 - 4.2 ANSI S3.22-1996

FONIX FP40 Portable Hearing Aid Analyzer 151Appendix J: Battery Simulator ImpedancesFP40 Battery Simulator Settings Battery Size Battery Chemi

Página 68

8 FONIX FP40 Portable Hearing Aid Analyzer Child Size Wedge Style Earhook Holds probe and reference microphones during real ear testing.Mounting Sl

Página 70 - 4.2.4 Viewing the results

Index 153 IndexAAccessories 6Optional 6Real-Ear 8Standard 5ACIC 55, 72Aided response 87ANSI 23, 43, 48ANSI S3.22-1987 55, 56, 58, 59, 60,

Página 71 - 4.3 ANSI S3.22-1987

154 FONIX FP40 Portable Hearing Aid Analyzer HHarmonic distortion 45Hearing aid setup 39body aid 41eyeglass aid 41ITE/ITC/CIC 39History of cha

Página 72 - 4.3.3 Viewing the results

Index 155 Shutdown 16Single frequency response 44, 107Smoothing 107Sound chamber 7Sound level calibrator 6Source type 21, 42Speaker, swing ar

Página 74

Introduction 9 Wedge Style Ear Hook Standard size Holds probe and reference microphones during real ear testing. Improved design eliminates need

Página 75 - 4.5 IEC

10 FONIX FP40 Portable Hearing Aid Analyzer 1.5 Layout & ControlsThis section gives a short explanation of the layout of the FP40 analyzer, it

Página 76

Introduction 11 1.5.2 Front Panel ButtonsThere are nine function key buttons in the top row of the FP40 front panel. The function of each of thes

Página 78 - 4.6 Profiler

12 FONIX FP40 Portable Hearing Aid Analyzer 1.5.4 Front Panel Knobs Amplitude Controls the amplitude or loudness of the test signal. Is also use

Página 79 - 4.6.3 Viewing the results

Introduction 13 1.5.6 Right Side Mounted Jack and ModuleLine input connector, IEC computer variety. Dual “snap in” fuse holder. Instrument will

Página 80 - 4.7 ACIC

14 FONIX FP40 Portable Hearing Aid Analyzer 1.6 FP40 SetupThis section describes how to set up the FP40 analyzer and prepare it for testing.1.6.1

Página 81

Introduction 15 If desired, plug in an external video monitor to the connector labeled “Video Monitor” on the back of the FP40. You can also plug

Página 82

16 FONIX FP40 Portable Hearing Aid Analyzer 1.7.3 Emergency ShutdownIf you find it impossible to turn off the instrument using the OPERATE button

Página 83

General Operation 17 Chapter 2: General OperationThis chapter discusses the general operation of the FP40 analyzer. You will learn how to navigate

Página 84

18 FONIX FP40 Portable Hearing Aid Analyzer 2.2.1 Making selections Notice the three columns in the General Setup Menu. Move the selection indi

Página 85 - Top View

General Operation 19 2.3 Using Function KeysThe front panel of the FP40 analyzer contains nine function keys, F1 through F9. These keys control th

Página 86

20 FONIX FP40 Portable Hearing Aid Analyzer Figure 2.3.1—The Main Coupler ScreenNotice that F1, F4, and F5 are written in large letters. They take

Página 87

General Operation 21 Note: If you toggle a function with SETUP F2 or SETUP F3, that function will remain in that set-ting until you explicitly chan

Página 88 - 5.2 The Target Screen

Table of ContentsChapter 1: Introduction1.1 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 89 - WARNING:

22 FONIX FP40 Portable Hearing Aid Analyzer 2.4.1.1 Pure-tone settingsThere are several different settings available in the General Setup Menu for

Página 90 - SANITATION NOTICE

General Operation 23 In determining the length of time needed for the proper measurement, a good rule is to use twice the published attack time of

Página 91 - 5.2.2 Creating a target

24 FONIX FP40 Portable Hearing Aid Analyzer Digital Speech ANSI uses the same speech spectrum as the Composite signal. This speech spectrum, taken

Página 92

General Operation 25 This type of distortion is only apparent when a composite signal source is used because pure-tone sweeps do not present more t

Página 93

26 FONIX FP40 Portable Hearing Aid Analyzer 2.5 Display ModeAlthough the FP40 comes standard with an LCD screen, it is easy to hook up an external

Página 94

General Operation 27 To turn on the battery current drain measurement in the Main Coupler Screen:1. Press F1, MENU, from the Main Coupler Screen.2.

Página 95

28 FONIX FP40 Portable Hearing Aid Analyzer 2.7 Using the DATA buttonAlthough it’s usually more useful to display test results in the familiar gra

Página 96 - 5.3.4 Viewing insertion gain

General Operation 29 To use an external sound chamber or an external speaker, plug it into the jack labeled “external speaker” on the back of the F

Página 97

30 FONIX FP40 Portable Hearing Aid Analyzer • If the print head is up, you will see the message PLEASE PUSH LEVER NEXT TO THE PRINTER. To correct

Página 98

General Operation 31 the External Printer Package from Frye Electronics containing a custom cable (PN 119-0312-00), a series-to-parallel converter,

Página 99

2.3.2 Customizing the function keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 202.4 Source Types . . . . . . . . . . .

Página 100 - 5.4.3 Viewing the SPL screen

32 FONIX FP40 Portable Hearing Aid Analyzer 2.10 The Opening ScreenThe Opening Screen of the FP40 (Figure 2.10) contains some useful information i

Página 101

General Operation 33 You may still be able to operate when the gauge shows empty, but you are on reserve. Battery oper-ation shuts down when the ba

Página 103 - 5.5.1 Viewing the AI display

Coupler Measurements 35 Chapter 3: Coupler MeasurementsThe Main Coupler Screen is the first operational screen you encounter when using the FP40.

Página 104 - 5.6 DSL Coupler Measurements

36 FONIX FP40 Portable Hearing Aid Analyzer 3.1.2 Viewing a Composite displayFigure 3.1.2—Composite coupler screen1. Display type: dB GAIN or dB

Página 105 - (I.D. 1.93 mm)

Coupler Measurements 37 altered since the analyzer was last leveled, invalidating the leveling. When in doubt, level the sound chamber again.3.2.1

Página 106

38 FONIX FP40 Portable Hearing Aid Analyzer 1. Set up the aid for testing as described in Section 3.3. If you are using a battery pill for test-in

Página 107 - WITH CUSTOM EARMOL

Coupler Measurements 39 3.3.1 Setting up a BTE1. Adjust the tone and gain controls of the aid to the desired positions. If desired, insert a batt

Página 108

40 FONIX FP40 Portable Hearing Aid Analyzer 3. Align the sound opening of the aid with the hole at the conical end of the coupler. Look through t

Página 109

Coupler Measurements 41 6. Slowly insert the microphone into the coupler. You may feel an initial resis-tance when the mic reaches the O-ring. Co

Página 110

3.4.6 Viewing harmonic distortion. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 453.4.7 Viewing battery current drain. .

Página 111

42 FONIX FP40 Portable Hearing Aid Analyzer 3.4 Frequency responsesThe Main Coupler Screen lets you take the frequency response of the hearing aid

Página 112

Coupler Measurements 43 Figure 3.4.1—Selecting a pure-tone sweepTo select a Composite/Digital Speech source:1. Press F1—MENU to enter the General

Página 113

44 FONIX FP40 Portable Hearing Aid Analyzer 3.4.3 Viewing multiple measurementsIt’s often useful to view several different frequencies responses t

Página 114

Coupler Measurements 45 2. Press F1 to enter the General Setup Menu.3. Select SOURCE with the AMPLITUDE and FREQUENCY knobs.4. Press START/STOP

Página 115 - 5.8 Miscellaneous

46 FONIX FP40 Portable Hearing Aid Analyzer 3. Use the FREQUENCY and AMPLITUDE knobs to select DISTORTION under PURE-TONE SETTINGS.4. Press START

Página 116 - 5.8.4 Data Display

Coupler Measurements 47 • You must position the reference microphone properly. If you don’t have the reference micro-phone positioned next to the

Página 117 - SIDE VIEW

48 FONIX FP40 Portable Hearing Aid Analyzer 3.5.1 Testing with Digital SpeechTo test digital aids with noise suppression, we have taken our standa

Página 118

Coupler Measurements 49 To use the Composite signal to test high end digital aids, you first need to figure out how long you can present the signal

Página 119

50 FONIX FP40 Portable Hearing Aid Analyzer 3.6.1 Preparing for the measurement1. Set up your analyzer so that the speaker is in the upright pos

Página 120

Coupler Measurements 51 5. Use the AMPLITUDE knob to select an appropriate source level. Typically this measurement should be done so that it is

Página 121

Chapter 5: Real-Ear Measurements5.1 Preparation for Real-Ear Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 735.1.1

Página 122 - 5.11 Testing Directional Aids

52 FONIX FP40 Portable Hearing Aid Analyzer 6. Press START/STOP once the measurement has stabilized. Compare the reverse measurement to the forwa

Página 123 - 5.11.2 Forward Measurement

Coupler Measurements 53 6. Press F2 or F3 (selected in step 3) to turn the CIC correction factors ON.7. Test as usual. See Section 3.4.2.See Figu

Página 124

54 FONIX FP40 Portable Hearing Aid Analyzer TYPE OF AID COUPLER COMMENT ITE, ITC MZ-1 BTE, or MZ-1 With custom earmold attached. (NOTE:

Página 125 - Chapter 6: Spectrum Analysis

Automated Test Sequences 55 Chapter 4: Automated Test SequencesThere are several different automated test sequences available on the FP40 analyze

Página 126

56 FONIX FP40 Portable Hearing Aid Analyzer Here are the major differences between ANSI 03 and ANSI 96:• When you set up an AGC aid for an ANSI 03

Página 127

Automated Test Sequences 57 To change the ANSI 03 settings:1. From the ANSI 03 screen, press F1 to enter the General Setup Menu.2. Use the AMPLI

Página 128

58 FONIX FP40 Portable Hearing Aid Analyzer 4.1.3 Running the test sequence1. Level the sound chamber if necessary. See Section 3.2 for details.2

Página 129 - Chapter 7: Telecoil Testing

Automated Test Sequences 59 8. SPA/HFA FOG: Three frequency average of a full-on gain measurement at 50 dB SPL9. REFTG TARGET and MEASURED: The

Página 130 - 7.2 Setup with the Telewand

60 FONIX FP40 Portable Hearing Aid Analyzer • Set the aid for the widest frequency response range.• Set AGC aids to achieve greatest possible com

Página 131

Automated Test Sequences 61 9. Use F5 to turn ON or OFF the telecoil measurement. The ANSI 96 telecoil measurement requires the Telewand.10. Pres

Página 132 - 7.4 Testing

5.8 Miscellaneous . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1075.8.1 Single frequency

Página 133 - Appendix A: Specifications

62 FONIX FP40 Portable Hearing Aid Analyzer 4.2.4 Viewing the resultsSee Figure 4.2.4 for an example of an ANSI 96 test sequence (including the te

Página 134

Automated Test Sequences 63 4.3 ANSI S3.22-1987The ANSI S3.22-1987 standard is an older version of the ANSI S3.22-1996 standard, discussed in Sec

Página 135 - SOUND CHAMBER

64 FONIX FP40 Portable Hearing Aid Analyzer 5. Press F7 to select the battery type used in the measurements.6. Press F8 to select whether or not

Página 136 - PHYSICAL DESCRIPTION

Automated Test Sequences 65 5. MAX: Maximum frequency response measured and the frequency at which it occurred6. SPA/HFA SSPL90: Three frequency

Página 137 - Appendix B: Calibration

66 FONIX FP40 Portable Hearing Aid Analyzer 4.4.1 Setting up for the testFollow the instructions found in Section 4.2.1 for instructions on settin

Página 138

Automated Test Sequences 67 4.4.3 Viewing the resultsSee Figure 4.4.3 for an an example of ANSI 92 results.1. NSPL90: RMS of Composite signal re

Página 139

68 FONIX FP40 Portable Hearing Aid Analyzer Follow the instructions from Section 3.3 to set up the aid in the test box for testing.4.5.2 Setting u

Página 140

Automated Test Sequences 69 4.5.3 Running the test sequence1. Level the sound chamber if necessary. See Section 3.2 for details.2. Set up the a

Página 141

Figure 4.5.4—IEC results4.6 ProfilerThe Profiler is an automated test sequence designed to tell you all the important information about the hearing a

Página 142

Automated Test Sequences 71 2. If you have chosen DSPA or DSPI in step 1, select the length of time each speech curve will be presented with F3.3

Página 143 - Appendix D: Custom RECD Test

viiiIf you are located in the European Union, please report all safety-related concerns to our autho-rized representative:Siemens Hearing InstrumentsA

Página 144

72 FONIX FP40 Portable Hearing Aid Analyzer Figure 4.6.3—Profiler results4.7 ACICThe ACIC test sequence is the ANSI 87 test sequence with CIC corr

Página 145 - The Basic Problem

Real-Ear Measurements 73 Chapter 5: Real-Ear MeasurementsThe Probe Option of the FONIX FP40 allows you to test hearing aids in the client’s ear, a

Página 146 - Acknowledgment

74 FONIX FP40 Portable Hearing Aid Analyzer Plug in the monitor headset if you want to be able to “listen in” on the sound received in the client’s

Página 147 - CIC CORRECTION FACTORS

Real-Ear Measurements 75 speaker postrecessedchannelspeaker jackFigure 5.1.1.2B—Upright sound field speakerNote: Do not close the chamber while it

Página 148

76 FONIX FP40 Portable Hearing Aid Analyzer 5.1.1.3 To set up an external sound field speakerIf you have a permanent location for your analyzer, y

Página 149

Real-Ear Measurements 77 Top ViewSpeaker(45 degrees)12"Reference MicProbe Mic12"Reference MicProbe MicFront ViewSpeaker(45 degrees)Figur

Página 150

78 FONIX FP40 Portable Hearing Aid Analyzer 5.1.2.3 To insert the probe tubeThere are several different methods used for properly inserting the pr

Página 151 - COUPLER PROBLEMS:

Real-Ear Measurements 79 Figure 5.1.2.3B – Unaided response with correct insertion of the probe tube5.1.2.4 To level the sound fieldLeveling is a

Página 152 - PRINTER PROBLEMS:

80 FONIX FP40 Portable Hearing Aid Analyzer Figure 5.1.2.4—Leveling the sound field speaker5.2 The Target ScreenThe Target screen is used to enter

Página 153

Real-Ear Measurements 81 WARNING:Choose OUTPUT LIMITING carefully (see procedure below). You do not want to damage your client’s hearing or cause

Página 154

Introduction 1 Chapter 1: Introduction1.1 Description Hearing aid analyzers are designed to give the user accurate information on how much amplif

Página 155

82 FONIX FP40 Portable Hearing Aid Analyzer SANITATION NOTICEDO NOT REUSE PROBE TUBES. Use a new probe tube for each ear to prevent the pos-sible s

Página 156

Real-Ear Measurements 83 Figure 5.2.1—The Target screen1. Insertion gain target graphical display.2. Fitting rule used in generation of the targ

Página 157 - Programming

84 FONIX FP40 Portable Hearing Aid Analyzer 6. Use the FREQUENCY and AMPLITUDE knobs to input the hearing threshold levels for the selected ear. T

Página 158 - & Hearing, 12(1): 47-54

Real-Ear Measurements 85 5.2.4 Creating your own targetAdvanced users may want to be able to input their own targets, without applying any partic

Página 159

86 FONIX FP40 Portable Hearing Aid Analyzer 5.3.1 Viewing the Insertion Gain screenHere is an picture of the Insertion Gain screen.1. Insertion g

Página 160

Real-Ear Measurements 87 4. Select the desired source with F7. Use either NORM (for a pure-tone sweep), or COMP (for a composite signal).5. Use

Página 161 - Index 153

88 FONIX FP40 Portable Hearing Aid Analyzer 6. Press F3 to select Aided 2, and repeat steps 3-5 to perform another measurement. A third aided resp

Página 162

Real-Ear Measurements 89 5.3.5 Testing Open Fit Hearing Aids Open fit hearing aids have been known to interfere with the reference microphone mea

Página 163 - Index 155

90 FONIX FP40 Portable Hearing Aid Analyzer 5.4 SPL Measurements (including real-ear DSL)The real-ear SPL screen permits the user to view all the m

Página 164

Real-Ear Measurements 91 5.4.2 Understanding the specifics of DSLThe Desired Sensation Level (DSL) method is a hearing aid selection and fitting

Modelos relacionados FONIX FP40

Comentários a estes Manuais

Sem comentários